High Pressure Gradient Controller
Model 601

Pump Unit Model 602
Pump Unit Model 603
The FLOM Gradient Controller with a pump unit achieves precise flow control and accurate solvent composition.
The flow volume, solvent composition, and flow rate can be programmed for precision control and automation
30 steps can be programmed with 20 files.
Each gradient method can be programmed up to 30 steps with 30 files.
The flow volume, solvent composition, and flow rate can be programmed during the solvent mixing process.
Pressure limit can be set at the program phase and the mixing phase.
The backlit LCD display shows flow rate and solvent composition as a graph for easy viewing.

Low Pressure Gradient Controller
Model 651
Capable of mixing 4 solvents.
Equipped with the pump component of Dual Pump 203, flow rate and flow volume are controlled for precision gradient.
Minimizes dead volume with pump's power valves.
The cooling system prevents air bubbles from forming within the tube.
Power voltage AC100V - 240V (2A), 50/60Hz
Ambience 10 - 30C, 20 - 80%RH (No condensation)
Number of solvents 4
Mixing method Precision valve proportioning
Gradient percentage range 0 - 100%
Flow rate range (ƒÓ3.2) 0.001 - 9.999ml/min in 0.001 ml/min steps
Flow rate range (ƒÓ7) 0.01 - 50.00 ml/min in 0.01 ml/min steps
Recommended flow rate range 0.1 - 9.99 ml/min. Mixer (optional) is recommended for 2.0 ml/min or higher
Pressure range 0 - 34.3MPa (0 - 350kgf/cm2)
Gradient profile Concave, convex, or linear.
Programming Steps: 1 - 99 steps X 99 programs
Analysis time: 999.9 min X 99 steps (0.1 min per step)
Display The 4-line-by-20-character backlid LCD digital display
(Erapsed time, solvent composition ratio, flow rate, pressure level, timer, etc.)
Chain program 1 - 99 steps
Purge solvent delivery system Yes
Cooling system Cooling system for vents to prevent air bubbling
Communication Inputs (contact closure or open contact): pump start, program start, reset button.
Outputs (contact closure): start output (at program start), error output (at error stop), end output (at the end of program/solvent flow).